7 set ipol - conditioning, Conditioning – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 724

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

708

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titrando, 855
Range

0 ... 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.4.5.7

SET Ipol - Conditioning

Tab: Method

SET Ipol Properties... Conditioning

Command name

Name of the command.

Input

25 characters

Conditioning

on | off (Default value: off)

Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift is
less than the entered start drift entered here. If this is the case, then the
status displayed in the Run subwindow in the Workplace program part
changes to COND READY.

Titrando, 855
Range

1 ... 999 µL/min

Default value

20 µL/min

Advertising