8.4 Manual Control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1528

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titrando, 855
Range

-125.0 ... 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Range

-127 ... 127 µA (Increment: 1)

Default value

5 µA

U(pol)

The polarization voltage is the potential applied to the polarized electrode
during an amperometric measurement.

Titrando, 855
Range

-1250 ... 1250 mV (Increment: 25)

Default value

400 mV

Titrino
Range

-1270 ... 1270 mV (Increment: 10)

Default value

400 mV

Reference temperature

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically conver-
ted to this reference temperature (usually 20 °C or 25 °C).

712
Range

-170.0 ... 500.0 °C

Default value

20.0

856
Range

-20.0 ... 150.0 °C

Default value

20.0

Selection

off

off
The conductivity at the measuring temperature is displayed.

Temperature compensation
Selection

Temperature coefficient | Sample solution

Default value

Temperature coefficient

Temperature coefficient
Manually enter a constant temperature coefficient.
Sample solution
Select a function for the temperature coefficient from the table Sample
solutions (TC conductivity)
.