4 meas ipol - evaluations, Evaluations – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 933

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

917

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

off
No stop.

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously.

Titrando, 855, 867
Range

-20.0 ... 150 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500 °C

Default value

25.0 °C

5.6.4.5.4

MEAS Ipol - Evaluations

Tab: Method

MEAS Ipol Properties... Evaluations

Command name

Name of the command.

Input

25 characters

Advertising