Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 960

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

944

■■■■■■■■

tiamo 2.2 (ProcessLab)

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the minimum measured
value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 °C/s

Default value

1.0 °C/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time and temperature for the
maximum measured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 °C/s

Default value

1.0 °C/s

Break-point evaluation

Note

This evaluation method is possible only with Titrando, 855, 856 and 867.

Break-point evaluation

on | off (Default value: off)

If this option is enabled, then sharp, almost 90° changes in directions in the
measurement curve will be evaluated.

EP criterion

Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value modification, even small changes in the

Advertising