1443

Min. beaker radius

Definition of the minimum radius which the beakers used on the rack must
have. If the lift is moved to the work position the beaker radii defined in
Rack table(see Chapter 6.11.2.1, page 1357) for general Sample posi-
tions
(see Chapter 6.11.3.2, page 1360) and Special beakers(see Chapter
6.11.3.4, page 1362)
will be compared with the Min. beaker radius. If
this minimum beaker radius is undercut then the run will be stopped and an
error message will appear. With off no check will be made.

Selection

1.0 … 100.0 mm | off

Default value

off

Lift rate

Lift speed for manual operation.

Range

3 ... 25 mm/s

Default value

25 mm/s

Axial distance

Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.

Range

100.0 ... 300.0 mm

814 USB Sample Processor XL
Selection

166.0 mm

Default value

166.0 mm

815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler, 864 Bal-
ance Sample Processor
Selection

196.0 mm

Default value

196.0 mm

Swing Head
This parameter is only shown when a swing head is mounted on the tower.

Serial number

Shows the serial number of the swing head.

Swing position

Lift position valid for all 4 external positions, at which the swing head turns
to the external positions.

Range

0 ... 235 mm

Default value

0 mm