Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 983

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

967

856
Range

0 ... 999999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Stop meas. value

Stops when the preset measured value has been reached since the start of
the measurement.

856
Range

0.0000 ... 2000.00 mS/cm

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:

Duration of measurement

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list. Max.
number of entries?

712
Range

0.4 ... 999999 s

Default value

2.0 s

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Advertising