3 dos - track call, 4 dos ph, 1 dos ph - overview – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1088: Dos ph

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1072

■■■■■■■■

tiamo 2.2 (ProcessLab)

Dosing rate in mL/min = dosing volume/(total duration - number of refillings
• filling time • 1/60)

5.6.6.3.3

DOS - Track call

Dialog window: Method

DOS pH/DOS U Monitoring [New]/[Properties] Track

call #

Monitoring

Selects the quantity at which a track is to be started when its limit value is
exceeded.

Selection

Measured value | Temperature | Any

Default value

Measured value

Any
Either one of the two sizes.

Limit exceeding

Selects the exceeded limit for which a track is to be started.

Selection

Lower limit | Upper limit | Any | OK

Default value

Any

Any
The track will be started if either the upper or the lower limit is exceeded.
OK
The track is started when the monitored quantities are once again within
the limit values (including hysteresis).

Track name

Selects the track that is to be started automatically.

Selection

Selection of available tracks

Note

If a track is called for that is already running then it will be allowed to
finish and then restarted when it is free again.

5.6.6.3.4

DOS pH

5.6.6.3.4.1

DOS pH - Overview

Dialog window: Method

DOS pH Properties... DOS pH - 'Command name'

Command for controlled dosing with measured value pH.

Devices
This command can be executed with the following devices:

Advertising