4 meas t/flow - evaluations, Evaluation – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 967

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

951

Stop measured value Temp.

Stops when the preset measured value has been reached since the start of
the measurement. off means no stop.

Range

50.0 ... 250.0 °C

Selection

off

Default value

off

5.6.4.8.4

MEAS T/Flow - Evaluations

Tab: Method

MEAS T/Flow Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves can
be activated and defined on this tab.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the minimum measured
value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 °C/s

Default value

1.0 °C/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the maximum measured
value will be interpolated from the measuring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 °C/s

Default value

1.0 °C/s