Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 929

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

913

I(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.

Titrando, 855, 867
Range

-125.0 ... 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Range

-127 ... 125.0 µA (Increment: 1)

Default value

5 µA

Electrode test

on | off (Default value: off)

If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement

Titrando, 855, 867
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.

Stirrer

Stirrer

Selection of the stirrer.

Titrando, 855, 867
Selection

1 | 2 | 3 | 4 | off