5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

956

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Sensor

Selection of an ISE sensor from the ISE sensors available in the sensor table.

Selection

'Sensor name' | ISE electrode

Default value

ISE electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.