Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1318

Advertising
background image

6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1302

■■■■■■■■

tiamo 2.2 (ProcessLab)

Input

24 characters

Default value

'empty'

Serial number

Serial number of dosing unit. With intelligent dosing units the serial number
is read off automatically and cannot be edited.

Input

10 characters

Default value

'empty'

Cylinder volume

Cylinder volume of dosing unit. With intelligent dosing units the cylinder
volume is read off automatically and cannot be edited. If in the method you
have selected a solution in a titration or dosing command then the cylinder
volume will be checked in the run.

Selection

2 mL | 5 mL | 10 mL | 20 mL | 50 mL

Default value

20 mL

Cylinder serial number

Serial number of cylinder. It is printed on new cylinders. With intelligent
dosing units it is read off automatically. It can be altered at any time, e.g. if
the cylinder has to be replaced.

Input

8 characters

Default value

'empty'

Parameters for preparation
Configures the parameters to be used in the commands PREP and
EMPTY.

Dosing port Prep/Empty

Dosing port via which the cylinder contents are to be ejected during prep-
aration and emptying.

Selection

Dosing port 1 | Dosing port 2 | Fill port | Special
port

Default value

Dosing port 1

Dosing rate Dosing port 1

Speed at which dosing is to be carried out at Dosing port 1. The maximum
dosing rate depends on the cylinder volume of the exchange unit used (see
below). When the function is carried out the speed will automatically be
reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum