5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1138

■■■■■■■■

tiamo 2.2 (ProcessLab)

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

774 Oven Sample Processor | 874 Oven Sample
Processor

Default value

874 Oven Sample Processor

Gas

Gas type

Gas type of the carrier gas.

Selection

Air | Nitrogen | other gas

Default value

Air

Gas flow factor

Factor for the correct determination of the gas flow. This parameter is only
visible for Gas type = other gas. The values for frequently used carrier gases
are listed in the device manuals.

Range

0.001 ... 9.999

Default value

1.000

Gas flow

Inlet

Selection of the inlet of the carrier gas

Selection

Pump | Valve

Default value

Pump

Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.

Flow rate

874
Range

10 ... 150 mL/min

Default value

50 mL/min