Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 935

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

919

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and temperature
for the maximum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

1.0 ... 2000.0 mV/s

Default value

25.0 mV/s

Break-point evaluation

Break-point evaluation

on | off (Default value: off)

If this option is enabled, then sharp, almost 90° changes in directions in the
measurement curve will be evaluated.

EP criterion

Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value modification, even small changes in the
measured value can be evaluated as a break point for a small measured value
range.

Range

0 ... 1.0

Default value

0.3

Slope

Minimum difference between the slope before and after the break point.
The smaller the difference, the more break-points will be found.

Range

0.0 ... 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Range

2 ... 20

Default value

5

Advertising