Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1371

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1355

c0 ... c4

Coefficients of the Chebyshev polynomial for the calculation of the function.

Range

1E-08 ... 1E+08

Default value

0.0

Assignment date

Display of the point in time at which the data for the calculation of the
temperature coefficient was entered or determined. This value is entered
automatically by tiamo.

Assignment method

Name of the method with which the automatic value assignment took place.
If the value has been entered manually then manual will be shown.

User

Short name of the user who was logged in during value assignment or who
entered the value manually. If work is not carried out using log in, then the
user logged in under Windows will be entered automatically.

Measuring input

Type and serial number of the measuring input at which the conductivity
sensor was connected.

Determination ID

Unambiguous identification for the determination with which the determi-
nation of the TC data was carried out.

Monitoring TC data
on | off (Default value: off)

If this option is switched on, then the period of validity of the data will be
monitored.

Validity

Period that the data is valid in days. If a value is entered here, then the date
in the field Next assignment will be adapted automatically.

Range

0 ... 999 Days

Default value

999 Days

Next assignment

Date on which the next assignment is to be carried out. The date can be

selected by pressing

in the dialog window Select date (see Chapter

2.5.1, page 83). After a date has been entered, the value for the Validity
will be automatically adapted.

Advertising