5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

483

greatest
Only the equivalence point with the greatest ERC value, i.e. the steepest
jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (U)

EP criterion

Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion will
not be recognized.

Titrino
Range

0 ... 200

Default value

5

[New]

Opens the dialog window Measured value # in which the parameters for
a new window can be entered (see Chapter 5.6.3.2.3.9, page 489).

[Properties]

Opens the dialog window Measured value # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.2.3.9, page 489).

[Delete]

Deletes the window selected in the table.

Evaluation with volume window (mL)

[New]

Opens the dialog window Volume window # in which the parameters for
a new window can be entered (see Chapter 5.6.3.2.3.10, page 490).

[Properties]

Opens the dialog window Volume window # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.2.3.10, page 490).

[Delete]

Deletes the window selected in the table.