Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 461

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

445

855
Selection

1 | 2 | 3

Default value

1

Titrino
Selection

internal D0

736, 751, 799
Selection

internal D0 | external D1 | external D2

Default value

internal D0

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Titrino
Selection

1 | 2 | diff.

Default value

1

Advertising