5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

982

■■■■■■■■

tiamo 2.2 (ProcessLab)

5.6.4.12.3

MEAS Ref - Measuring parameters

Tab: Method

MEAS Ref Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Start wavelength

Lower limit of the spectrum.

Range

200.0 ... 1100.0 nm

Default value

400.0 nm

End wavelength

Upper limit of the spectrum.

Range

200.0 ... 1100.0 nm

Default value

1000.0 nm

Integration time

Integration time for spectrum recording.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Range

0 ... 100 Pixels

Default value

0 Pixels