4 analog out, 1 analog out - overview, Analog out – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1218

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1202

■■■■■■■■

tiamo 2.2 (ProcessLab)

5.6.11.4

ANALOG OUT

5.6.11.4.1

ANALOG OUT - Overview

Dialog window: Method

ANALOG OUT Properties... ANALOG OUT - 'Command

name'

Devices
This command can be executed with the IO Controller.

Appearance
The command has the following appearance:

Mode of operation
The values which are written to the analog outputs can be either calculation
results or fix values.

The values, the set values and unit of all of the analog outputs used in the
command are saved to command variables.

Parameters
The parameters for the command ANALOG OUT are configured on the
following 2 tabs:

General

Analog outputs

Command variables
The following command variables are generated in the method run by the
command ANALOG OUT and can be used in formulas under the designa-
tion 'Command name.Variable designation':

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has not
ever been started

.DEF

Set value which is defined in the configuration for an IO port.

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Command
has never been ended

.UNI

Mapping unit which is defined in the configuration for an IO
port.

Advertising