5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1039

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.

Titrando, 855, 867
Range

0 ... 999999 s

Default value

10 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:

Titrando, 855, 867
Range

0 ... 999999 s

Default value

110 s

Titrino
Range

0 ... 9999 s

Default value

110 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s (Increment: 0.08)

Default value

2.0 s

Temperature

Temperature

The temperature at which calibration is carried out can be entered manually.
If a temperature sensor is connected and the Temperature measure-
ment
is set on the General/Hardware tab under Sensor to automatic
or continuous, then the temperature will be measured continuously. The
value is used for temperature compensation with the pH measurement.