5.6.3.5.2.10

KFT Ipol - Fix endpoint evaluation

Dialog window: Method

KFT Ipol Properties... Additional evaluations Fix

endpoint evaluation

[New]/[Property] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.5.3

KFT Upol

5.6.3.5.3.1

KFT Upol - Overview

Dialog window: Method

KFT Upol Properties... KFT Upol - 'Command name'

Command for Karl Fischer Titrations with amperometric measurement
(selectable polarization voltage).

Devices
This command can be executed with the following devices.

Titrando: 835, 836, 841, 852, 857, 890, 901, 906, 907

Titrino: 720, 736, 751, 758, 784, 795, 799

Appearance
The command has the following appearance:

without conditioning

with conditioning