6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1304

■■■■■■■■

tiamo 2.2 (ProcessLab)

Tubing parameters
Defines the length and diameter of the tubing connected to the dosing unit.
The port assignment can also be altered. These parameters are important
for carrying out the dosing unit commands PREP and EMPTY correctly, as
the volumes of the tubing connections have to be taken into account.

Note

Default values have already been entered for the tubing parameters; these
correspond to the dimensions of the standard tubing supplied. If you do
not make any alterations to the tubing connections then you do not need
to modify the tubing parameters. For information on length and diameter
of other tubings see Products/Accessories and columns under

http://

products.metrohm.com

.

Dosing port 1

Port

Port to be used as Dosing port 1.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 1

Length

Length of tubing attached to Dosing port 1.

Range

0.0 ... 999.9 cm

Default value

40.0 cm

Diameter

Diameter of tubing attached to Dosing port 1.

Range

0.0 ... 9.9 mm

Default value

2.0 mm

Dosing port 2

Port

Port to be used as Dosing port 2.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 3

Length

Length of tubing attached to Dosing port 2.