8 Manual Control

tiamo 2.2 (ProcessLab)

■■■■■■■■

1535

For Type = Stirrer:

0: Duration of the switch-on pulse fixed at 500 ms.

>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in 10
ms
Example:
Parameter 1
= 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on 225
ms, switched off 75 ms

Range

-214783647 ... 214783647

Default value

0

Status

Display of the current status.

Selection

ACTIVE | INACTIVE | ERROR

[Start]

Sets the digital output to ACTIVE.

[Stop]

Sets the digital output to INACTIVE.

Timer control

Port description

Display of the port description.

Port

Display of the internal description of the port.

Type

Display of the port type.

Parameter 1

The meaning of Parameter 1 depends on the Type of the digital output that
has been set.

When Type = Stirrer: stirrer output in %, pulse duration dependent on
Parameter 2.

Range

-214783647 ... 214783647

Default value

100

Parameter 2

The meaning of Parameter 2 depends on the Type of the digital output that
has been set.