Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 820

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

804

■■■■■■■■

tiamo 2.2 (ProcessLab)

Conditioning

on | off (Default value: on)

Enable/Disable the conditioning. If the conditioning is disabled, the following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift is
less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series
tab will then switch to COND READY.

Range

1 ... 999 µg/min

Default value

20 µg/min

Drift correction

If this parameter is set to automatic or manual the drift correction ( = Drift
value
× Time for drift correction) caused by empty consumption during
titration is deducted from the total consumption. The Time for drift cor-
rection
is the period between end of conditioning and end of titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Range

0.0 ... 999.9 µg/min

Default value

0.0 µg/min

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Range

1 ... 999999 s

Selection

off

Default value

off