5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1042

■■■■■■■■

tiamo 2.2 (ProcessLab)

3 and multiple point calibration
With at least 3 calibration solutions the calibration function is determined
according to the method of least squares. The sought calibration param-
eters s, E(0) and c

B

are calculated iteratively. If the value for c

B

is lower

than aprox. 0.2 % of the lowest concentration, it is set equal to zero. In
such a case a linear relation between U

i

and log(c

i

) results.

The variance Var is calculated and output in any case with at least 4
calibration standards. If c

B

= 0, the variance is even calculated for 3

standards.

U

i

: measured potential

U

i,c

: potential calculated out of the calibration curve

n : number of measuring points
F : degrees of freedom (2 for c

B

= 0, 3 for c

B

≠ 0)

Command variables
The following command variables are generated in the method run by the
command CAL LOOP Conc and can be used in formulas under the desig-
nation 'Command name.Variable designation':

Identifi-
cation

Description

.BLV

Blank value calculated from the calibration

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started

.ENP

Electrode zero point calculated from the calibration

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command has
never ended

.LCO

Loop counter = current quantity of completed runs, both of
Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

.SLO

Electrode slope calculated from the calibration