Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1252

Advertising
background image

6.2 Administration

■■■■■■■■■■■■■■■■■■■■■■

1236

■■■■■■■■

tiamo 2.2 (ProcessLab)

gency stop, which are always possible for all running determinations and
series.

If this check box is deactivated then the newly logged-in user will have all
the rights for the opened workplaces that his/her group assignment author-
izes him/her to. For running determinations the new user name applies
immediately (Audit Trail entries, Save, Export, etc.).

Selection

on | off

Default value

off
FDA default: off

Automatic logout after

If this check box is activated then the user will be logged out automatically
when no operating functions have been carried out with the keyboard or
the mouse within this time. After this automatic logout, only a member of
the same user group as the one to which the previously logged-off user
belongs or a user with administrator rights can log in again.

Selection

on | off

Default value

off
FDA default: off

Range

1 ... 60 min

Default value

10 min
FDA default: 10 min

Note

The immediate termination of all determinations on all activated work-
places with the emergency stop button still remains possible after an
automatic logout.

Password policies

Enforce unique password

If this check box is activated then this ensures that a user can use a given
password only once.

Selection

on | off

Default value

off
FDA default: on

Enforce use of special characters

If this check box is activated then this ensures that the password must
contain at least one special character (@, #, ~ etc.).

Advertising