5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1014

■■■■■■■■

tiamo 2.2 (ProcessLab)

Number of additions

Number of times the standard solution is added.

Range

1 ... 19

Default value

3

Volume auxiliary solution

Volume of the auxiliary solution (e.g. buffer) which was added to the sample
before the first standard addition.

Range

0.0000 ... 9999.999 mL

Default value

0.0000 mL

Stop volume

Volume of auxiliary solution (e.g. buffer) that is to be added to the sample
before the first standard addition.

Range

0.0 ... 9999.9 mL

Default value

100.0 mL

Addition solution

Note

Displayed only for Solution = not defined.

Concentration

Concentration of addition solution.

Range

0.001 ... 999999.999

Default value

1.000

Device

Concentration unit of addition solution.

Selection

mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L

Default value

ppm

Control parameters

Dosing rate

Speed at which the volume increments are added.

Selection

slow | medium | fast

Default value

fast