2 result - monitoring, Monitoring – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1164

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1148

■■■■■■■■

tiamo 2.2 (ProcessLab)

[Save as template]

Open the dialog window Save result template in order to save the result
parameters as template for the creation of new results (see Chapter
5.6.8.2.6.2, page 1152)
.

5.6.8.2.5.2

Result - Monitoring

Tab: Method

CALC Properties... [New] [Continue] Result - Overview

The parameters for monitoring the result are defined on this tab.

Result monitoring

Result monitoring

on | off (Default value: off)

The question of whether the result is within the defined limit values is moni-
tored while calculating the result during the determination if this option is
enabled.

Lower limit

Lower limit for the result.

Input

10 digits

Upper limit

Upper limit for the result.

Input

10 digits

Message

Message

The message defined here can be put out to different targets if the upper
limit is exceeded or if the lower limit is not reached. The text editor for

entering or changing the message is opened with

or by double-clicking

into the text field (see Chapter 2.5.2, page 84). The formula editor can also
be called up inside the text field (see Chapter 2.4, page 25).

Input

Text (unlimited)

Display message

on | off (Default value: on)

If this option is enabled, all active tracks are stopped and the message
defined in the text box will be displayed in the event that the limits are
exceeded. You can select whether you want to continue the stopped tracks
with [Continue] or - if defined - to execute the Action listed below.

Advertising