Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1065

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1049

Parameters
The parameters for the command CAL Cond are configured on the follow-
ing 3 tabs:

General/Hardware
Parameters for devices, sensors and stirrers.

Measuring parameters
Parameters for setting the measurement procedure.

Conductivity standard
Defining the values or the type of the input for calibration with one
standard.

Command variables
The following command variables are generated in the method run by the
command CAL Cond and can be used in formulas under the designation
'Command name.Variable designation':

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.CLC

Cell constants determined by calibration for the sensor
used in the command with conductivity sensors (only
MEAS Cond and CAL Cond)

.DBL

Total duration for the processing of the command in s

.EME

Final measured value (measured value after processing of
the command) in the unit of the measured value

.ETE

End temperature (temperature following processing of
the command) in °C for the last measurement

.FIN

Command status: 1 = Command has been ended at
least once
, 0 = Command was started but not
ended
, invalid = Command was not started (varia-
ble not available)
.

.IME

Initial measured value (measured value at the start of the
command) in the unit of the measured value for the last
measurement

.ITE

Initial temperature (temperature before processing of the
command) in °C for the last measurement

.LP.MEA

Measured value for the last measuring point in the meas-
uring point list in the unit of the measured value