5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

428

■■■■■■■■

tiamo 2.2 (ProcessLab)

Note

The name of the assigned method variables is displayed automatically for
the sample data variables ID1 … ID16, Sample size, Unit, Sample
position
in the subwindow Run on the tab Single determination and
in the subwindow Information on the tab Sample.

Fixed value

on | off (Default value: off)

If this option is enabled, then the method variables can be assigned a fixed
value. For variables of the type Date/Time, the date must be entered in the
dialog window 2.5.1Select date. No formulas can be entered in this field.

Type = Number
Range

-1.0E-99 ... 1.0E+99

Type = Text
Input

100 characters

Type = Date/Time
Selection

'Date'

Check at start

on | off (Default value: on)

If this option is enabled, the start test will check if the method variable is
valid and if the limit values are maintained when monitoring is switched on.

Note

Switching off this option can be useful if the method variables are still
invalid at the start of the method and will not be defined until there is a
REQUEST command during the run.

Comment

Freely selectable comment on the method variable.

Input

1000 characters