5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

874

■■■■■■■■

tiamo 2.2 (ProcessLab)

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.9.9

TET - Endpoint

Dialog window: Method

TET Properties... Thermometric evaluation [New]/

[Properties]

Endpoint

Reaction type

Selection of the reaction type according to which the curve is to be evalu-
ated.

Selection

exothermic | endothermic

Default value

exothermic

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.

for reaction type = exothermic
Range

-999999 ... 0

Default value

-5

for reaction type = endothermic
Range

0 ... 999999

Default value

5

5.6.3.9.10

TET - Fixed endpoint evaluation

Dialog window: Method

TET Properties... Additional evaluations Fixed end-

point evaluation #

Response

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Range

0.0 ... 60.0 °C