6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1361

Rack offset

The rack offset is a production-dependent tolerance value between the
upper part of the rack and the lower part. The value is determined by a rack
adjustment and displayed here. If necessary, it can be edited.

Range

-5.00 ... 5.00 °

6.11.3.3

Properties - Lift positions

Tab: Configuration

Rack data [Edit] Properties... Rack data Lift positions

Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions except
those that are defined as Special beaker.

Work position

Work position for Lift 1. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.

Range

0 ... 235 mm

Rinse position

Rinse position for Lift 1. This lift position is used for rinsing the electrodes.

Range

0 ... 235 mm

Shift position

Shift position for Lift 1. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be selected
so that a safe movement across the entire rack is possible at any time.

Range

0 ... 235 mm

Special position

Special position for Lift 1. This additional definable position can be used e.g.
during pipetting so that the tip is just immersed in the sample solution.

Range

0 ... 235 mm

Tower 2
Defines the lift positions for Tower 2. These apply for all rack positions except
those that are defined as Special beaker.

Work position

Work position for Lift 2. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.