Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1309

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1293

Rel. standard dev. (s rel)

Relative standard deviation for the titer determination in %.

Monitoring titer validity
on | off (Default value: off)

If this option is switched on then the validity of the titer will be monitored.

Validity

Validity of the titer in days. If a value is entered here then the date for the
Next titer determination will be automatically adapted.

Range

0 ... 999 Days

Default value

999 Days

Next titer determination

Date on which the next titer determination is to be carried out. The date can

be selected by pressing

in the dialog window Select date (see Chapter

2.5.1, page 83). After the date has been entered, the Validity field will be
adapted automatically.

Set value

    Date Titer det.+ 999 days

Message
The message options can only be edited if monitoring is switched on.

Message by E-mail

on | off (Default value: off)

The message is additionally sent to the E-mail address defined under [E-
mail]
if this option is activated. The message is sent in text format.

[E-mail]

[E-Mail] opens the Send E-mail window (see Chapter 2.6.1, page 86).

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above

Action
The actions can only be edited if monitoring is switched on.

If during titer monitoring it is found that the validity period of the titer has
expired then one of the following actions will be triggered automatically at
Start test.