5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

828

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

off

Default value

off

Titrino
Range

0.01 ... 150.00 mL/min

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.8.6.7

STAT pH - Monitoring

Tab: Method

STAT pH Properties... Monitoring

Command name

Name of the command.

Input

25 characters

The following monitoring processes can be switched on and defined on this
tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is switched on then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.

Lower limit pH

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.