5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

649

or after a waiting period. The volume that has been added when the end-
point is reached gives the calculable reagent consumption.

Reagent addition takes place in three phases during the titration:

Initial dosing
During this phase the dosing rate increases continuously. It starts with
the Min. rate and increases to the Max. rate.

Continuous dosing
In this phase addition is carried out at the Max. rate until the Control
range
is reached.

Dynamics
Dosing is finely controlled in this range. Shortly before the endpoint is
reached addition is only carried out at the Min. rate (Control range).