5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

763

Titrando
Selection

1 | 2 | 3 | 4 | off

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos.

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command. This parameter is displayed only for 751, 785, 784, 795 and
799.

5.6.3.5.3.3

KFT Upol - Start conditions

Tab: Method

KFT Upol Properties... Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the titration
is started.

Initial measured value

Note

Is displayed only for Titrando.

The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters: