5 tet - stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 885

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

869

Range

1 ... 200

Default value

50

Damping until

Range from the start of the measurement in which the curve is additionally
damped. This parameter allows to minimize the influence of titration arti-
facts (e.g. spikes), which occur at the beginning of the measurement and
can mistakenly be interpreted as endpoints. The damping functions as a
linear ramp, i.e. its influence is the highest at the beginning and then decrea-
ses linearly to the defined volume.

Range

0.0 ... 50.0 mL

Default value

0.2 mL

5.6.3.9.5

TET - Stop conditions

Tab: Method

TET Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Range

0.01 ... 50.00 mL

Selection

off

Default value

off

Stop measured value

Stops when the measured value entered for a measuring point has been
reached since the start of the titration. Thereby, the direction of the meas-
urement curve (ascending or descending), which is determined after record-
ing the first 50 measuring points, is taken into account.

Range

0.0 ... 60.0 °C

Selection

off

Default value

off

Stop time

Stops when the preset time (including start conditions) has elapsed since the
start of the titration.