10 kft upol - fix endpoint evaluation – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 792

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

776

■■■■■■■■

tiamo 2.2 (ProcessLab)

[Delete]

Deletes the calculated measured value selected in the table.

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can be
taken over and saved from existing measuring commands (i.e. MEAS) which
can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.5.3.10

KFT Upol - Fix endpoint evaluation

Dialog window: Method

KFT Upol Properties... Additional evaluations Fix

endpoint evaluation

[New]/[Property] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-200.0 ... 200.0 µA

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL