4 stdadd man - measuring parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1015

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

999

Addition solution

Concentration

Concentration of addition solution.

Range

0.001 ... 999999.999

Default value

1.000

Device

Concentration unit of addition solution.

Selection

mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L

Default value

ppm

5.6.4.15.4

STDADD man - Measuring parameters

Tab: Method

STDADD man Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement with drift control
The measurement is stopped as soon as the Signal drift or the Stop
measured value
is reached or a defined Waiting time has elapsed.

Measured value drift

The measured value is only accepted if the signal drift defined here has been
fallen short of.

Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time

Advertising