5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

781

Sensor

Sensor

Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

Polarization current at the indicator electrode.

Selection

2 µA | 5 µA | 10 µA | 20 µA | 30 µA

Default value

10 µA

Note

When changing this parameter, make sure that the polarization current
of the indicator electrode, the endpoint and the control range are linked
closely with one another (see Chapter 5.6.3.6.4, page 783).

Electrode test

on | off (Default value: on)

If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

851, 852
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.