5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

624

■■■■■■■■

tiamo 2.2 (ProcessLab)

Additional measured values
Definition of additional measured values of other measuring commands
which are saved as additional columns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command MET Upol and can be used in formulas under the designation
'Command name.Variable designation':

Identification

Description

.BP{x}.DME

Measured value differential for the break point × (1…
9)

.BP{x}.MEA

Measured value for the break point × (1…9) in the unit
of the measured value

.BP{x}.TEM

Temperature for the break point × (1…9) in °C

.BP{x}.TIM

Time for the break point × (1…9) in s

.BP{x}.VOL

Volume for the break point × (1…9) in mL

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used
for the command

.DBL

Total duration for the processing of the command in s

.DSC

Time for processing all start conditions s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.EP{x}.DME

Measured value differential for the endpoint × (1…9)

.EP{x}.MEA

Measured value for the endpoint × (1…9) in the unit
of the measured value

.EP{x}.MEP

Quantity of endpoints in the window × (1…9); 1 = 1
endpoint
, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift

.EP{x}.TEM

Temperature for the endpoint × (1…9) in °C

.EP{x}.TIM

Time in s until the endpoint × (1…9) is reached

.EP{x}.VOL

Volume for the endpoint × (1…9) in mL