Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 837

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

821

off
The measured vales will not be applied until after the maximum waiting
time has passed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s

Pause 1

Pause 1

Waiting time, e.g. until the electrode has settled down, before a start volume
is added.

Titrando, 855
Range

0 ... 999999 s

Default value

0 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate indi-
cated.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Titrino
Range

0.00 ... 999.99 mL

Default value

0.00 mL