3 digital in, 1 digital in - overview, Digital in – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1215

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1199

>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in 10
ms
Example:
Parameter 1
= 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on 225
ms, switched off 75 ms

Range

-214783647 ... 214783647

Default value

0

5.6.11.3

DIGITAL IN

5.6.11.3.1

DIGITAL IN - Overview

Dialog window: Method

DIGITAL IN Properties... DIGITAL IN - 'Command

name'

Devices
This command can be executed with the IO Controller and with the Stepping
Motor Controller.

Appearance
The command has the following appearance:

Mode of operation
The statuses of all digital inputs will be entered and saved in command var-
iables at the time of the command (syntax Command name.Variable
designation
)

Parameters
The parameters for the command DIGITAL IN are configured on the fol-
lowing 2 tabs:

General
Parameters for device.

Stop criteria
Parameters for defining the stop criteria.

Command variables
The following command variables are generated in the method run by the
command DIGITAL IN and can be used in formulas under the designation
'Command name.Variable designation':