7 met ipol - additional evaluations, Additional evaluations – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 631

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

615

Evaluation with volume window (mL)

[New]

Opens the dialog window Volume window # in which the parameters for
a new window can be entered (see Chapter 5.6.3.3.4.10, page 621).

[Properties]

Opens the dialog window Volume window # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.4.10, page
621)
.

[Delete]

Deletes the window selected in the table.

5.6.3.3.4.7

MET Ipol - Additional evaluations

Tab: Method

MET Ipol Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab:

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value
, time or volume) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fix
endpoint (see Chapter 5.6.3.3.4.11, page 622).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fix end-
point selected in the table (see Chapter 5.6.3.3.4.11, page 622).