6 met ipol - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 629

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

613

5.6.3.3.4.6

MET Ipol - Potentiometric evaluation

Tab: Method

MET Ipol Properties... Potentiometric evaluation

Command name

Name of the command.

Input

25 characters

Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of the
following three evaluation methods can be chosen here:

Potentiometric evaluation

Selection

Evaluation without window | Evaluation with
measured value window (Upol)
| Evaluation
with volume window (mL)

Default value

Evaluation without window

Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition
will be applied across the entire range of the titration curve.
Evaluation with measured value window (Upol)
With this option set, up to 9 regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within these
windows and additionally meet the parameters defined for each window
will be recognized. Per window only one EP will be recognized. The
defined windows with their parameters are shown in the window table
and can be edited with the following buttons:
Evaluation with volume window (mL)
This option is visible only for Titrando and 855.. With this option up to
9 regions (windows) can be defined on the volume axis. Only those
equivalence points which lie within these windows and additionally meet
the parameters defined for each window will be recognized. Only one
EP will be recognized per window. The defined windows with their
parameters are shown in the window table and can be edited with the
following buttons:

Evaluation without window

EP criterion

Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion will
not be recognized.

Range

1 ... 999 mV

Default value

30 mV