background image

6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1320

■■■■■■■■

tiamo 2.2 (ProcessLab)

Document message
The message that the validity of the calibration has expired will be auto-
matically saved with the determination.
Display message
A message is displayed and you can select whether you want to continue
the run or cancel it. If the run is continued then the message that the
validity of the calibration has expired will be saved automatically with the
determination.
Cancel determination
The running determination will be automatically canceled. The following
message must be confirmed with [OK].

6.7.3.4

Properties - Limits

Tab: Configuration

Sensors Edit Sensor Limits

Note

This tab is displayed only for sensors of the type pH electrode, ISE elec-
trode
or conductivity sensor.

Slope monitoring (for pH and ISE electrodes only)
on | off (Default value: off)

If this option is enabled then a check will be made to see whether the
Slope determined during calibration is within the defined limits. If it is out-
side them a corresponding message appears and you can select whether
you will nevertheless accept the value or whether the previous value is to be
retained.

Lower limit

Lower limit value for the slope.

for pH electrode
Range

-2.147E+9 ... 2.147E+9 %

Default value

96.0 %

for ISE (+1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

55.0 mV

for ISE (-1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-61.0 mV