5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

752

■■■■■■■■

tiamo 2.2 (ProcessLab)

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Range

0.0 ... 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size of
the titration cell in order to prevent any overflow.

Titrando
Range

0.000000 ... 9999.99 mL

Default value

20,000 mL

Selection

off

Default value

off

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Titrando
Range

1 ... 999999 s

Selection

off

Default value

off

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Range

0 ... 999 s

Default value

0 s

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Range

0 ... 99 s

Default value

0 s

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the Event
message
Start titration is sent to the command with a SEND command.