2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

tiamo 2.2 (ProcessLab)

Variable name

Description

Data
type

SEN

Indication whether the end of the sample table has been reached; 1 = yes, 0 =
no

Number

SIN

Indication whether the determination has been started as a single determination or
within a series; 1 = single determination, 0 = series determination

Number

SLI

Sample table actual line (number)

Number

STA

Indication whether the statistics is activated; 1 = yes, 0 = no

Number

STC

Start counter

Number

USN

Short name of the logged-in user

Text

ORG

Method run: 1 = original determination, 0 = reprocessing

Number

STO

Indication whether the determination has been stopped (manual stop, stop via SEND
command, emergency stop) or terminated normally ; 1 = stopped, 0 = termina-
ted normally

Number

2.4.3.8

Common variables

Dialog window: Formula Editor

Variables

Common variables are global variables, which are adopted from the corre-
sponding table of the program part Configuration, where the common
variables can be defined, at the start of the determination and assigned to
the determination. The Variables field of the Formula editor lists all Com-
mon Variables
which are available for the current method, sorted accord-
ing to name.

Syntax
'CV.Variable name.Variable name'

Examples: 'CV.TestDate', 'CV.TestTime.VAL', 'CV.Average-
Temp.UNI'

You can select the common variables directly in the formula editor under
Variables/Common Variables in order to avoid syntax errors.

Available Common Variables

Variable name

Description

Data
type

VAL

Value of the Common Variable (facultative, i.e. 'CV.Test.VAL' = 'CV.Test')

Text,
number
or date/
time

UNI

Unit of the Common Variable

Text