Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 652

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

636

■■■■■■■■

tiamo 2.2 (ProcessLab)

as the measured value alterations between two measuring points are then
of the same order of magnitude as the noise.

Titrando, 855
Range

0.0001 ... 9.9999 mL

Default value

0.05 (slow), 0.10 (optimal), 0.20 (fast) mL

Titrino
Range

0.001 ... 9.999 mL

Default value

0.05 (slow), 0.10 (optimal), 0.20 (fast) mL

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C