5 det ipol - stop conditions, Stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 522

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

506

■■■■■■■■

tiamo 2.2 (ProcessLab)

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

5.6.3.2.4.5

DET Ipol - Stop conditions

Tab: Method

DET Ipol Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the titration are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Advertising