tiamo 2.2 (ProcessLab)

■■■■■■■■

567

Gran evaluation

Note

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accordance
with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 878).

Procedure

Selection of the Gran procedure.

Selection

Normalized | Standard

Default value

Normalized

Initial volume

Volume which must be present in the measuring vessel before the command
is started.

Range

0.01 ... 9999.99 mL

Default value

50.0 mL

Lower limit pH

Value for the lower limit of the Gran evaluation range.

Range

-20.000 ... 20.000

Default value

-20.000

Upper limit pH

Value for the upper limit of the Gran evaluation range.

Range

-20.000 ... 20.000

Default value

20.000

5.6.3.3.2.8

MET pH - Additional measured values

Tab: Method

MET pH Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings: